Product overview
The SYN5606 type time relay tester is a high precision multi function time relay testing equipment designed and developed by Xi'an synchronous Electronic Technology Co., Ltd.. This product is based on the JJF 1282-2011 "electronic time relay calibration specification" and the JJF 1400-2013 "time relay tester calibration specification" requirements of the production of a multi-function, high reliability, special precision time measuring instrument.The high stability constant temperature crystal oscillator is used as the time benchmark, and the 7 inch LCD touch screen has AC and DC voltage output. It uses FPGA technology of large scale integrated circuit and digital control to realize high precision time interval test. The whole machine has the advantages of high stability and high accuracy, perfect function, convenient operation and strong anti-interference ability.
Product function
1) switching between internal and external frequency standards.
2) electronic time relay test;
3) power on / off / turn on / off delay electronic time relay test;
4) large touch screen for real-time display.
5) directly output AC and DC voltage and AC voltage control port.
6) output the result directly to the computer through the serial port.
Product characteristics
A) highly integrated and highly precise.
B) good stability and reliable performance.
C) liquid crystal display panel is simple and intuitive.
Typical application
1) metrological calibration departments and research institutes.
2) the instrument can be widely applied to all kinds of time relay tests.
3) single and dual channel time interval measurement applications.
Technical index
Input signal | External reference | Routes | 1 way |
frequency | 10MHz | ||
wave form | sine | ||
level | 5dBm~13dBm | ||
Physical interface | BNC | ||
Time relay measurement | Scope of application | Power on / off / on / off | |
measuring range | 0.001s~9999.999s | ||
Accuracy | Better than + (1 x 10-7 x T0 + 0.5ms) | ||
Physical interface | BNC | ||
Time interval measurement | Scope of application | Single channel / dual channel | |
Trigger level | -5V to +5V (set) | ||
measuring range | 30ns ~ 99999.999999 990s | ||
Accuracy | Better than + (1 x 10-7 x T0 + 30ns) | ||
Physical interface | BNC | ||
output signal | DC voltage output | 10-200V adjustable | 1 way |
AC voltage output | 220V | 1 way | |
ocxo | Routes | 1 way | |
frequency | 10MHz | ||
wave form | sine | ||
level | More than 7dBm | ||
Daily aging rate | Less than 5 x 10-10/ days | ||
Second stability | Less than 5 x 10-11/s | ||
Accuracy | Less than 1 x 10-7 | ||
Preheating time | More than 12 hours | ||
Physical interface | BNC | ||
RS232C serial port | Routes | 1 way | |
level | RS232C | ||
Serial port format | Measurement results and setting status | ||
Physical interface | DB9 | ||
External AC power control interface | Routes | 1 way | |
AC voltage | 10-250V | ||
Environmental characteristics | working temperature | 0 C ~ + 50 C | |
relative humidity | Less than 90% (40 C) | ||
Storage temperature | -30 C ~ + 70 C | ||
Power supply | AC 220V + 10%, 50Hz + 5%, power less than 30W | ||
Chassis dimensions | 3U, 19 "standard chassis (upper rack) 482mm (wide) X300 (deep) x150mm (high) | ||
Option | Customized products according to customer requirements. |