Product overview
The SYN5619 type phase noise analyzer is a high-precision noise tester designed by Allen synchronous Electronic Technology Co., Ltd., which is designed and developed by ourselves. The phase noise and the Allen variance are simultaneously tested.Using advanced phase measurement technology, the use of 9 Inch Touch screen, 4 core processor, 4G memory, 4G display, 60G solid state drive and industrial keyboard, etc., the switch will take about ten seconds or less within a few seconds to test results, and achieve graphical user interface.Industrial keyboard and touchscreen can be used at the same time. The operation is simple. Just press the start to test in seconds. Without skilled engineers, you can easily learn and use.
The phase noise analyzer can display the measurement results in real time, and the results are accurate and reliable. No additional data processing, such as Allen variance calculation, is needed.The cost of accurate phase noise and Allen variance measurement is significantly reduced, covering almost all commonly used frequency sources.Achieving -140 dBc/Hz@1 Hz makes it the best choice for analyzing ultra low phase noise frequency source.
Key words: frequency stability test, phase noise tester, phase noise analyzer
Product function
1) phase noise and stability test for 500kHz-30MHz frequency simultaneously.
Product characteristics
A) simultaneously testing phase noise and stability.
B) supports input and reference tests with different frequencies;
C) highly integrated, high precision, ultra low background;
D) fast and stable, reliable performance.
E) one button operation, LCD panel, concise and intuitive;
F) a cost-effective solution.
G) no measurement calibration requirements, save time.
Typical application
1) phased array radar, radio navigation system, automatic control system and other tests.
2) scientific research institutes and other occasions requiring high quality requirements for frequency signals.
3) in the metrology sector, high-precision oscillator equipment manufacturers or users.
Technical index
Reference and measured signals | Routes | 1 roads | |
Physical interface | BNC | ||
frequency range | 500kHz ~ 30MHz | ||
Input amplitude | -5dBm to +20dBm (recommended +5dBm to +15dBm) | ||
Input impedance | 50 Omega | ||
Maximum DC voltage | + 5V | ||
Frequency offset analysis range | 1Hz ~ 100kHz | ||
Clutter level (5MHz) | Less than -100dBc (less than -120dBc (typical value)) | ||
Allen variance (10MHz@0.5Hz BW) | Less than 1E-13/1S (5E-14/1s (typical value)) Less than 5E-15/1000S (1E-15/1000s (typical value)) | ||
phase noise | 1Hz, @5MHz<-140dBc (<-145dBc (typical value)) 1Hz, @25MHz<-130dBc (<-135dBc (typical value)) 10kHz, @5MHz<-170dBc (<-175dBc (typical value)) 10kHz, @25MHz<-165dBc (<-165dBc (typical value)) | ||
Preheating time | No less than 30 minutes. | ||
Environmental characteristics | working temperature | 15 centigrade to + thirty five degrees centigrade. | |
relative humidity | Less than 90% (40 C) | ||
Storage temperature | -20 C ~ + 50 C | ||
Power supply | AC 220V + 10%, 50Hz + 5%, power less than 30W | ||
Chassis dimensions | 4U, 19 "standard chassis (upper rack) 482mm (wide) X500 (deep) x178mm (high) | ||
Option | Customized products according to customer requirements. |
Guide for selection of related products
Model | Input signal | Testing function | characteristic | Other |
SYN5104 type time integrated parameter tester | GPS Beidou | Time standard source, time difference measurement, time accuracy, frequency accuracy, message accuracy, Zhou Bo measurement, temperature and humidity measurement, time recording | Most time frequency signals can be tested. | 4U, 19 "(upper rack) AC220V, 30W |
Model SYN5106 high precision clock tester | GPS Beidou | Time standard source, time difference measurement, time accuracy, frequency accuracy, timing error measurement | Clock test of electric energy meter | 4U, 19 "(upper rack) AC220V, 30W |
SYN5301 millisecond meter time calibrator | 10MHz reference | Verification of mechanical stopwatch, electronic stopwatch, pointer type electric stopwatch, millisecond meter, digital electric stopwatch and so on, combined with 4 functions of stopwatch calibrator, day difference meter / calibrator, pointer type electric stopwatch calibrator, standard time interval generator and so on. | High accuracy and powerful function | 3U, 19 "(upper rack) AC220V, 30W |
SYN5301A type stopwatch verification fixture | Control electrical signal | Verification of mechanical stopwatch, electronic stopwatch, and various kinds of timer and other fixtures. | Flexible and convenient, high automation. | Module (360x110x140mm) |
SYN5301C precision photoelectric converter | One precision laser source, one high-precision photodetector, two sets of brackets. | High precision | AC220V power adapter | |
SYN5602 calibrator for electronic parking time charge device | GPS Beidou | At the same time, 1~18 electronic parking meter (parking meter) or parking area electronic time charge device can be checked at the same time. | The targets are all superior to the requirements of the national metrological verification regulation of JJG1010-2013 electronic parking timed tariff meter. | Portable suitcase AC 220V + 10% |
SYN5603 GPS time (event) recorder | GPS | 3 pulse time recording | High accuracy | 1U, 19 "(upper rack) AC 220V + 10% |
SYN5605 multi channel time interval measuring instrument | Reference 10MHz | Simultaneous measurement of 36 pulse signals | High precision, multi-channel simultaneous measurement | 3U, 19 "(upper rack) AC 220V + 10% |
SYN5605A digital time interval measuring instrument | Reference 10MHz | Measuring the time difference between 1 or 2 pulse signals | High precision time interval measurement | 3U, 19 "(upper rack) AC 220V + 10% |
SYN5605C type time correlated single photon counting system | Reference 10MHz | Simultaneous measurement of multiple pulses | PS magnitude accuracy | 3U, 19 "(upper rack) AC 220V + 10% |
SYN5607 type phase meter | The phase difference is measured at 0 degrees to 360 degrees or + 180 degrees. | Accuracy 0.1 degree | 3U, 19 "(upper rack) AC 220V + 10% | |
SYN5608 digital phase standard | Output two adjustable sine wave with high precision phase, ultra wide voltage range and ultra low distortion. | Accuracy 0.1 degree | 3U, 19 "(upper rack) AC 220V + 10% | |
Type SYN5610 pulse generator | Reference 10MHz | Single channel, dual channel output | High precision | 4U, 19 "standard chassis. |
SYN5619 phase noise analyzer | Phase noise and stability test for 500kHz-30MHz frequency simultaneously | Less than 1E-13/1S (5E-14/1s (typical value)) 1Hz@5MHz<-140dBc (<-145dBc (typical value)) | 4U, 19 "standard chassis. | |
Option | GPS Beidou dual-mode receiver, constant temperature crystal oscillator OCXO, rubidium clock, lightning arrester, customized antenna cable (50 meters, 80 meters, 100 meters, etc.), expand the output path, PTP measurement, NTP measurement, 4U standard chassis, network photoelectric conversion kit, according to customer requirements customized products.Customize products with wide temperature range, and make similar products according to customer's requirements. |